Direct thromboaspiration efficacy for mechanical thrombectomy is related to the angle of interaction between the aspiration catheter and the clot.

Gianmarco BernavaAndrea RosiJosé BotoOlivier BrinaZsolt KulcsarChristoph CzarnetzkiEmmanuel CarreraKarl SchallerKarl-Olof LovbladPaolo Machi
Published in: Journal of neurointerventional surgery (2019)
In our series, an angle of interaction between the aspiration catheter and the clot of ≥125.5° was significantly associated with successful clot removal. The prediction of the angle of interaction on pretreatment imaging may help operators to select the most adequate mechanical thrombectomy technique on a case by case basis.