Real-time monitoring of atmospheric ammonia based on modifier-enhanced vacuum ultraviolet photoionization ion mobility spectrometry.
Xueying BaiShimin LiuWei HuangWeiguo WangDongming LiAibo WangYi ChenYuanzhi ZhangHuaiwen CangHaiyang LiPublished in: Analytical methods : advancing methods and applications (2023)
Ammonia (NH 3 ) plays an important role in the atmospheric environment such as the formation of PM 2.5 , the concentration monitoring of which could hence help in the air quality assessment. In this study, a method for quantitative monitoring of atmospheric NH 3 was developed based on modifier-enhanced selectivity detection using a homemade vacuum ultraviolet photoionization ion mobility spectrometry (VUV-PI-IMS). To enhance the resolution and sensitivity of measuring NH 3 , 2-butanone as the gas modifier was introduced into the drift tube with the drift gas. Atmospheric NH 3 can be selectively detected, where the peak-to-peak resolution ( R P-P ) of 7.69 was obtained. The product ions were identified to be [C 4 H 8 O] 2 NH 4 + by using a homemade time-of-flight mass spectrometer. The calculated limit of detection (LOD) was 0.39 ppbv improving about 10 times. For the most common concentration variation of atmospheric NH 3 in the range of 10-100 ppbv, the linear curve was obtained with R 2 of 0.997. Lastly, the VUV-PI-IMS was used to monitor the evolution of atmospheric NH 3 near our laboratory and mounted on a car for monitoring the regional distribution of atmospheric NH 3 in Dalian, China. The results also showed that VUV-PI-IMS has a promising application prospect in monitoring the concentration of atmospheric NH 3 and supporting the air quality assessment.