Quantitative effects of bilirubin structural photoisomers on the measurement of direct bilirubin via the vanadate oxidation method.
Makoto AriokaKosuke KoyanoYasuhiro NakaoMiyo OzakiShinji NakamuraHiroyuki KiuchiHitoshi OkadaSusumu ItohKoji MuraoTakashi KusakaPublished in: Annals of clinical biochemistry (2023)
Under ambient WLED light, serum sample generated 1.3 mg/dL BSIs in 1 h. Approximately 34% (0.44 mg/dL) of BSI concentrations was measured as DB when using the vanadate oxidation method according to the above equation.