Incidence of Pseudotumors in a Dual Modular Stem Construct With and Without Metal-on-Metal Bearing Surface.

Tahsin M RahmanNicholas B FrischBrian DarrithIshan PatelCraig D Silverton
Published in: The Journal of the American Academy of Orthopaedic Surgeons (2021)
The incidence of pseudotumor formation in MoM total hip arthroplasties was more than five times the incidence associated with non-MoM bearings. The correlation between serum titanium concentration and pseudotumor size suggests that titanium levels may be a useful indicator for pseudotumor formation in patients with this particular titanium modular neck femoral implant. No notable correlation was observed between serum cobalt and chromium concentration and pseudotumor formation or size.
Keyphrases
  • total hip
  • risk factors
  • total knee arthroplasty