Computational FEM Model and Phantom Validation of Microwave Ablation for Segmental Microcalcifications in Breasts Using a Coaxial Double-Slot Antenna.

Kristian Segura FélixGeshel D Guerrero LópezMario F J Cepeda RubioJosé Irving Hernández JácquezFrancisco G Flores GarcíaArturo Vera HernándezLorenzo Leija SalasEva C Orozco Ruiz de la Peña
Published in: BioMed research international (2021)
The SWR was obtained by triplicate experiments in the phantom, and the measurements had a difference of 0.191 between the minimum and maximum SWR values, implying a change of power reflection of 0.8%. The average of the three measurements was compared with the simulation that did not consider blood perfusion. The comparison yielded a change of 0.104, representing a 0.2% change in power reflection. Discussion. Both experimentation in phantom and simulations demonstrated that ablation therapy can be performed using this antenna. However, an additional optimization procedure is warranted to increase the efficiency of the antenna.