Where can you wear your Libre? Using the FreeStyle Libre continuous glucose monitor on alternative sites.
Rosemary M HallSophie DyhrbergArthur McTavishLindsay McTavishBrian CorleyJeremy D KrebsPublished in: Diabetes, obesity & metabolism (2022)
The FSL-CGM is clinically accurate when the sensors are applied to alternate chest or back sites. Greater variability occurs during rapid changes in glucose concentration with all sensor sites compared with venous glucose. Understanding these variabilities allows appropriate use of an economically viable continuous glucose monitor.