Comparison between Abbott m2000 RealTime and Alinity m STI systems for detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Mycoplasma genitalium.

Björn HerrmannKarin Malm
Published in: European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology (2021)
The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis, 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genitalium, 22 (7.9%) of 279 samples were positive in the Alinity assay and 18 (6.5%) in the m2000 assay, for which DNA extraction was performed on an m2000sp instrument combined with in-house real-time PCR. The Alinity assay has at least the same sensitivity as the m2000 assay. The specificity was evaluated by discrepancy analysis.