Dual-angle open field wavefront sensor for simultaneous measurements of the central and peripheral human eye.

Dmitry RomashchenkoLinda Lundström
Published in: Biomedical optics express (2020)
We have developed a novel dual-angle open field wavefront sensor. This device captures real-time foveal and peripheral Zernike aberrations, while providing natural binocular viewing conditions for the subjects. The simultaneous data recording enables accurate analysis of changes in ocular optics with accommodation overcoming any uncertainties caused by accommodative lag or lead. The instrument will be used in myopia research to study central and peripheral ocular optics during near work and to investigate the effects of optical myopia control interventions. Proof of concept measurements, performed on an artificial eye model and on 3 volunteers, showed good repeatability with foveal-peripheral data synchronization of 65 msec or better. The deviations from subjective cycloplegic refractions were not more than 0.31 D. Furthermore, we tested the dual-angle wavefront sensor in two novel measurement schemes: (1) focusing on a close target, and (2) accommodation step change.