Login / Signup

Validation of Heart Rate Extracted From Wrist-Based Photoplethysmography in the Perioperative Setting: Prospective Observational Study.

Eveline H J MestromRuben DeneerAlberto G BonomiJenny MargaritoJos GelissenReinder HaakmaHendrikus H M KorstenVolkher ScharnhorstArthur R Bouwman
Published in: JMIR cardio (2021)
This study shows that an OHRM equipped with a PPG sensor can measure HR within the ECG reference standard of -5 bpm to 5 bpm or -10% to 10% in the perioperative setting when the PPG signal is of sufficient quality. This implies that an OHRM can be considered clinically acceptable for HR monitoring in low acuity hospitalized patients.
Keyphrases
  • heart rate
  • heart rate variability
  • blood pressure
  • cardiac surgery
  • patients undergoing
  • quality improvement