Development and optimisation of grid inserts for a preclinical radiotherapy system and corresponding Monte Carlo beam simulations.
Marcus FiskPejman RowshanfarzadDavid PfefferléMatthew Fernandez de VianaJulian CabreraMartin Andrew EbertPublished in: Physics in medicine and biology (2024)
Objective . To develop a physical grid collimator compatible with the X-RAD preclinical radiotherapy system and create a corresponding Monte Carlo (MC) model. Approach . This work presents a methodology for the fabrication of a grid collimator designed for utilisation on the X-RAD preclinical radiotherapy system. Additionally, a MC simulation of the grid is developed, which is compatible with the X-RAD treatment planning system. The grid was manufactured by casting a low melting point alloy, cerrobend, into a silicone mould. The silicone was moulded around a 3D-printed replica of the grid, enabling the production of diverging holes with precise radii and spacing. A MC simulation was conducted on an equivalent 3D grid model and validated using 11 layers of GAFChromic EBT-3 film interspersed in a 3D-printed water-equivalent phantom. A 3D dose distribution was constructed from the film layers, enabling a direct comparison with the MC Simulation. Main results . The film and the MC dose distribution demonstrated a gamma passing rate of 99% for a 1%, 0.5 mm criteria with a 10% threshold applied. The peak-to-valley dose ratio and output factor at the surface were determined to be 20.4 and 0.79, respectively. Significance . The pairing of the grid collimator with a MC simulation can significantly enhance the practicality of grid therapy on the X-RAD. This combination enables further exploration of the biological implications of grid therapy, supported by a knowledge of the complex dose distributions. Moreover, this methodology can be adapted for use in other systems and scenarios.