General Algorithm Applicability in Determining DBS Lead Orientation: Adapting 2D and 3D X-Ray Techniques for SenSightTM Leads.
Stefan HunscheDieter FeddersAlexandra HellerbachMarkus EichnerJochen WirthsTill A DembekVeerle Visser-VandewalleHarald TreuerPublished in: Stereotactic and functional neurosurgery (2024)
The 2D and 3D algorithms, initially formulated for discerning the circular orientation of a C-shaped marker, were adapted to the marker of the Medtronic SenSight™ lead. Statistical analyses revealed a significant level of agreement between the two methods. Our findings highlight the adaptability of these algorithms to different markers, achievable through both low-dose intraoperative 2D X-ray imaging and standard CT imaging.