Login / Signup

Ultra-fast LC-MS/MS in therapeutic drug monitoring: Quantification of clozapine and norclozapine in human plasma.

Lewis CouchmanDanielle S FisherKrithika SubramaniamSimon A HandleyRobert J BoughtflowerChristopher M BentonRobert J Flanagan
Published in: Drug testing and analysis (2017)
A novel approach to high-throughput, targeted liquid chromatography-tandem mass spectrometry (LC-MS/MS) analysis has been developed. A single chromatographic system can be used for the analysis of a range of 20 drugs and metabolites with a total analysis time of 36 s (one 96-well plate of prepared samples per hour). To demonstrate the applicability of this approach to quantitative analysis, a method has been validated for the therapeutic drug monitoring of clozapine and norclozapine following automated extraction from human plasma. Chromatographic retention times were 11.4 and 12.4 s for norclozapine and clozapine, respectively (for both analytes the chromatographic peak width was less than 1 s). Comparison with a conventional LC-MS/MS method (5 min analysis time) showed excellent agreement. This new approach offers analysis times more akin to flow-injection analysis, but is likely to be more widely applicable because of chromatographic resolution from residual matrix components and isobaric interferences.