Dried Blood Spots and Miniaturized Ultrasonic Nebulization Microplasma Optical Emission Spectrometry for Point-of-Care Testing of Blood Lithium.

Yuanyuan LiJie TangYunhan XiaoTian RenJiahui YangYao LinChengbin Zheng
Published in: Analytical chemistry (2024)
Despite the significant importance of blood lithium (Li) detection in the treatment of bipolar disorder (BD), its point-of-care testing (POCT) remains a great challenge due to tedious sample preparation and the use of large-footprint atomic spectrometers. Herein, a system coupling dried blood spots (DBS) with a point discharge optical emission spectrometer equipped with a miniaturized ultrasonic nebulizer (MUN-μPD-OES) was developed for POCT of blood Li. Three microliters of whole blood were used to prepare a dried blood spot on a piece of filter paper to which 10 μL of eluent (1% (v/v) formic acid and 0.05% (v/v) Triton-X) was added. Subsequently, the paper was placed onto the vibrating steel membrane of the ultrasonic nebulizer and powered on to generate aerosol. The aerosol was directly introduced to the μPD-OES for quantification of Li by monitoring its atomic emission line at 670.8 nm. The proposed method minimized matrix interference caused by high levels of salts and protein. It is worth noting that the MUN suitably matches the needs of DBS sampling and can provide aerosolized introduction of Li into the assembled μPD-OES, thus eliminating all tedious sample preparation and the need for a commercial atomic spectrometer. Calibration response is linear in the therapeutic range and a limit of detection (LOD) of 1.3 μg L -1 is well below the Li minimum therapeutic concentration (2800 μg L -1 ). Li in mouse blood was successfully detected in real-time using MUN-μPD-OES after intraperitoneal injection of lithium carbonate, confirming that the system holds great potential for POCT of blood Li for patients with BD.
Keyphrases