Gastric artery embolization: studying the effects of catheter type and injection method on microsphere distributions within a benchtop arterial model.

Shaphan R JerniganJason A OsborneGregory D Buckner
Published in: Biomedical engineering online (2020)
Catheter type had a significant effect on targeting accuracy across all injection methods. The Endobar catheter exhibited a higher targeting accuracy in pairwise comparisons with the other two injection catheters across all injection schemes and when considering the Endobar catheter with the manifold injection method vs. each of the catheters with the manual injection method; the differences were significant in three of four analyses. The injection method effect was not statistically significant across all catheter types and when considering the Endobar catheter/Endobar manifold combination vs. Endobar catheter injections with manual and pressure-replicated methods.