Login / Signup

Computed Tomography Attenuation of Three-Dimensional (3D) Printing Materials-Depository to Aid in Constructing 3D-Printed Phantoms.

Yuktesh KalidindiAravinda Krishna GanapathyYash R NayakAnusha ElumalaiDavid Z ChenGrace BishopAdrian SanchezBrian AlbersAnup S ShettyDavid H Ballard
Published in: Micromachines (2023)
Three-dimensionally printed phantoms are increasingly used in medical imaging and research due to their cost-effectiveness and customizability, offering valuable alternatives to commercial phantoms. The purpose of this study was to assess the computed tomography (CT) attenuation characteristics of 27 resin materials from Formlabs, a 3D printing equipment and materials manufacturer. Cube phantoms (both solid and hollow constructions) produced with each resin were subjected to CT scanning under varying tube current-time products with attenuation measurements recorded in Hounsfield units (HU). The resins exhibited a wide range of attenuation values (-3.33 to 2666.27 HU), closely mimicking a range of human tissues, from fluids to dense bone structures. The resins also demonstrated consistent attenuation regardless of changes in the tube current. The CT attenuation analysis of FormLabs resins produced an archive of radiological imaging characteristics of photopolymers that can be utilized to construct more accurate tissue mimicking medical phantoms and improve the evaluation of imaging device performance.
Keyphrases