Comparison of retinal vessel diameter measurements from swept-source OCT angiography and adaptive optics ophthalmoscope.
Xinwen YaoMengyuan KeYijie HoEmily LinDamon W K WongBingyao TanLeopold SchmettererJacqueline ChuaPublished in: The British journal of ophthalmology (2020)
The inter-rater agreement was excellent for vessel diameter measurement in both OCTA protocols (ICC=0.92) and AOO (ICC=0.98). The measured vessel diameter was widest from the OCTA 3×3 mm2 (55.2±16.3 µm), followed by OCTA 9×9 mm2 (54.7±14.3 µm) and narrowest by the AOO (50.5±15.6 µm; p<0.001). Measurements obtained from both OCTA protocols were significantly wider than the AOO scan (OCTA 3×3 mm2: mean difference Δ=4.7 µm, p<0.001; OCTA 9×9 mm2: Δ=4.2 µm, p<0.001). For vessels >45 µm, it appeared to be larger in OCTA 3×3 mm2 scan than the 9×9 mm2 scan (Δ=1.9 µm; p=0.005), while vessels <45 µm appeared smaller in OCTA 3×3 mm2 scan (Δ=-1.3 µm; p=0.009) CONCLUSIONS: The diameter of retinal vessels measured from OCTA scans were generally wider than that obtained from AOO scans. Different OCTA scan protocols may affect the vessel diameter measurements. This needs to be considered when OCTA measures such as vessel density are calculated.