To date, thermistors are used to continuously monitor the body temperature of newborn babies in the neonatal intensive care unit. The thermistor probe is attached to the body with a strong adhesive tape to ensure that the probe stays in place. However, these strong adhesives are shown to increase microbial growth and cause serious skin injuries via epidermal stripping. The latter compromises the skin's ability to serve as a protective barrier leading to increase in water loss and further microbial infections. In this article a new approach is introduced that eliminates the need for an adhesive. Instead, two kinds of fluorophores are entrapped in a skin friendly chitosan gel that can be easily wiped on and off of the skin, and has antimicrobial properties as well. A CCD camera is used to detect the temperature dependent fluorescence of the fluorophore, tris(1,10-phenthroline)ruthenium(II) while 8-aminopyrene-1,3,6-trisulfonic acid serves as the reference. This temperature sensor was found to have a resolution of at least 0.13°C.
Keyphrases