A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit.
M F Iparraguirre NuñoveroLucila PiaseckiA V K SegatoV P D WestphalenU X Silva NetoEverdan CarneiroPublished in: International endodontic journal (2021)
Compared with alginate, the conductive gel of the ProTrain kit negatively affected most of the electronic readings. The three evaluated ERCLMDs had similar accuracy in locating the apical foramen when using the mark 0.0/APEX and alginate as the embedding media. Although the ProTrain platform allows radiographs to be exposed in both bucco-lingual and proximal views, no difference was found between the views when evaluating the apical limit provided by the 0.5 mark in mandibular premolars embedded in alginate.