Minimalist transcatheter aortic valve replacement misses paravalvular regurgitation: Incidence and echocardiographic distribution of missed paravalvular regurgitation.

Syed ZaidAhmed Abu HaniyehAkiva RosenzveigAaqib MalikJoshua B GoldbergCenap UndemirDaniel SpevackSteven L LansmanGilbert H L TangHasan Ahmad
Published in: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2022)
Intraoperative TTE-guided minimalist TAVR either misses nearly half of ≥mild PVR or underestimates PVR by ≥1 grade when compared to predischarge TTE. Posterior PVR jets are more likely to be missed. Transesophageal echo guidance may help minimize missing PVR. Further studies are warranted.