Global and peak local specific absorption rate control on parallel transmit systems using k-means SAR compression model.
Xianglun MaoSumra BariDavid J LoveJoseph V RispoliPublished in: Magnetic resonance in medicine (2020)
The developed SAR compression model is advantageous for pTx framework given the narrower overestimation bound and control over the compression ratio. Results also illustrate that a moderate increase of maximum RF power can be useful for reducing the maximum local SAR deposition.
Keyphrases