Login / Signup

H-D Analysis Employing Energy Transfer from Metastable Excited-State He in Double-Pulse LIBS with Low-Pressure He Gas.

M PardedeT J LieJ IqbalM BilalR HedwigM RamliA KhumaeniW S BudiN IdrisS N AbdulmadjidA M MarpaungI KarnadiI TanraZ S LieH SuyantoD P KurniawanKoo Hendrik KurniawanK KagawaM O Tjia
Published in: Analytical chemistry (2018)
A laser-induced-breakdown-spectroscopy (LIBS) experiment with a unique double-pulse setup and operated in low-pressure (3 kPa) He ambient gas is performed to study the detection of light elements, such as hydrogen (H) and deuterium (D), as well as elements of high excitation energies, such as fluorine (F) and chlorine (Cl), which are usually difficult to detect using ordinary LIBS techniques. A nanosecond Nd:YAG laser operated in its fundamental wavelength with energy of 54 mJ is focused onto the Al target to generate the He plasma. Another picosecond Nd:YAG laser operated in its fundamental wavelength with energy of 2 mJ is focused onto the sample surface and activated 2 μs before the operation of the nanosecond laser. The application to polyvinyl chloride (PVC) and polytetrafluoroethylene (PTFE) samples produces sharp and high-intensity Cl- and F-emission lines. Meanwhile, the sharp and well-resolved H-D-emission lines with merely 0.18 nm wavelength separation are also clearly detected from a zircaloy sample. Further measurement of a set of zircaloy samples containing different concentrations of D yields a linear calibration curve with a zero intercept. The detection limit of D is found to be about 10 ppm.