FT-ICR MS Spectral Improvement of Dissolved Organic Matter by the Absorption Mode: A Comparison of the Electrospray Ionization in Positive-Ion and Negative-Ion Modes.
Qing-Long FuChao ChenYang LiuManabu FujiiPingqing FuPublished in: Analytical chemistry (2023)
Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS) in the absorption mode has a superior performance over the conventional magnitude mode. However, this improved performance for the analysis of dissolved organic matter (DOM) in negative-ion and positive-ion modes of electrospray ionization [ESI(-) and ESI(+), respectively] remains unknown. This study systemically compared the improved performance by the absorption mode for DOM FT-ICR MS spectra acquired with the low-field and high-field magnet instruments between two charge modes. The absorption mode enhanced the resolution and signal-to-noise ratio values of DOM peaks with factors of 1.88-1.94 and 1.60-1.72, respectively. The significantly higher improvement of mass resolution for the ESI(+) mode than that for the ESI(-) mode could resolve the extensive occurrence of mass doublets in the ESI(+) mode, yielding some formulas exclusively identified in the ESI(+) mode. The findings of this study have systemically demonstrated the superiority of the absorption mode in improving the spectra quality during the routine FT-ICR MS postdata analysis and highlighted its great potential in characterizing the molecular composition of DOM using the FT-ICR MS technique in both ESI(-) and ESI(+) modes.