Login / Signup

Development a stacking pad design for enhancing the sensitivity of lateral flow immunoassay.

Tsung-Ting TsaiTse-Hao HuangChung-An ChenNatalie Yi-Ju HoYi-Ju ChouChien-Fu Chen
Published in: Scientific reports (2018)
Lateral flow immunoassays (LFIAs) have wide application in point-of-care testing, particularly in resource-poor settings. To achieve signal amplification in a gold nanoparticle-based lateral flow assay without an additional procedure or the need for complex fabrication, a new and simple method was developed for using a "stacking pad" configuration that adds an additional membrane between the conjugation pad and test pad to the conventional AuNP-based LFIA format. This design helps to accumulate the antibody and antigen on the stacking pad, hence extending the antigen/antibody binding interactions to enhance the test's detection sensitivity. With the enhanced lateral flow assay, as low as 1 ng/mL of Protein A and 15.5 ng/mL of C-reactive protein can be visualized with the naked eye. We also successfully applied the stacking pad system in the analysis of C-reactive protein in human serum and synovial fluid samples. These results suggest that this stacking pad LFIA can provide sensitive and on-site prognosis for detection in synovial fluid and serum samples in resource-limited settings.
Keyphrases
  • label free
  • high throughput
  • single cell
  • small molecule
  • transcription factor
  • nucleic acid
  • dna binding