Comparative analysis of two optical biometry devices: high wavelength swept source OCT versus partial coherence interferometry.

Eszter SzalaiAdrienne Csutak
Published in: International ophthalmology (2021)
Our results implicated an overall good reproducibility of anterior keratometry, AL, ACD and WTW measurements for IOLMaster and ANTERION. The discrepancies between their measurements resulted in significant difference in the calculated IOL power for SRK/T, Hoffer and Holladay formulas, but not for Haigis formula.
Keyphrases
  • optical coherence tomography
  • high speed
  • diabetic retinopathy
  • optic nerve
  • preterm infants