Application of a Carbon Fiber Microelectrode as a Sensor for Apocynin Electroanalysis.
Slawomir MichalkiewiczAgata SkorupaMagdalena JakubczykKarolina BębaczPublished in: Materials (Basel, Switzerland) (2024)
In this study, a carbon fiber microelectrode (CF) was applied for the investigation of the electrochemical behavior of the natural antioxidant, apocynin (APO). Given the limited solubility of APO in water, a mixture of anhydrous acetic acid (AcH) with 20%, v / v acetonitrile (AN) and 0.1 mol L -1 sodium acetate (AcNa) was used. The electrochemical properties of APO were examined through linear sweep voltammetry (LSV), differential pulse voltammetry (DPV), and cyclic voltammetry (CV). The anodic oxidation of APO, which is the basis of the method used, proved to be diffusion-controlled and proceeded with a two-electron and one proton exchange. Both radicals and radical cations, arising from the first and second step of electrode reactions, respectively, underwent subsequent chemical transformations to yield more stable final products ( E q C i E i C i mechanism). Using optimized DPV conditions, the anodic peak current of APO at a potential of 0.925 V vs. Ag/AgCl showed a good linear response within the concentration range of 2.7 × 10 -6 -2.6 × 10 -4 mol L -1 . The detection and quantification limits were determined as 8.9 × 10 -7 and 2.7 × 10 -6 mol L -1 , respectively. The developed DPV method enabled the successful determination of APO in herbal extracts and in dietary supplements. It should be noted that this is the first method to be used for voltammetric determination of APO.