Login / Signup

Dual catheter and double-lumen cerebrospinal fluid shunt systems with backflow mechanisms.

Anup PatelDavid QiJacqueline BoyleMartin MorrisJulian Jwchun Lin
Published in: Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery (2023)
The preliminary results of connecting a standard proximal catheter in series with a double-lumen catheter show a slight change in the percent occlusion from the baseline status several times when the retrograde flow occurred via one limb of the catheter. Additionally, the de-obstruction seems related to the length of the interval of the backflow and the initial percentage occlusion of the proximal catheter. The statistical analysis does not reveal a statistically significant reduction in occlusion in the proximal catheter with either backflow interval.
Keyphrases
  • ultrasound guided
  • cerebrospinal fluid
  • genome wide
  • dna methylation
  • coronary artery