EP radiofrequency generators: Significant offsets between selected and delivered power?

Felix BourierBernhard SchwarzAmir BrkicLara WolffVerena SemmlerMarc KottmaierElena RisseMarta TelishevskaStephanie BrooksSarah LengauerMarielouise KornmayerKatharina Koch-BüttnerFlorian BergerGabriele HesslingIsabel DeisenhoferTilko Reents
Published in: Journal of cardiovascular electrophysiology (2017)
The Ampere® and SmartAblate® generator models deliver accurate power as selected by the user. The power delivered by the EP Shuttle® generator exceeds the selected power by up to 40% depending on impedance. The findings were confirmed in ex vivo porcine heart experiments and should be considered in clinical practice.