ST waveform analysis for monitoring hypoxic distress in fetal sheep after prolonged umbilical cord occlusion.

Peter AndriessenAlex ZwanenburgJudith O E H van LaarRik VullingsBen J M HermansHendrik J NiemarktReint K JellemaDaan R M G OpheldersTim G A M WolfsBoris W KramerTammo Delhaas
Published in: PloS one (2018)
The STAN device showed limitations in detecting hypoxemia in fetal sheep after prolonged UCO. The STAN device produced high false positive event rates during baseline and did not detect T/QRS changes adequately after prolonged fetal hypoxemia. During 14% of baseline and 62% of the UCO period, the STAN-device could not process the ECG signal, despite its good quality. Resolving these issues may improve the clinical performance of the STAN device.
Keyphrases
  • umbilical cord
  • mesenchymal stem cells
  • heart rate variability
  • left ventricular
  • data analysis