Finite Element Modeling of Residual Hearing after Cochlear Implant Surgery in Chinchillas.

Nicholas CastleJunfeng LiangMatthew SmithBrett PetersenCayman MatsonTara EldridgeKe ZhangChung-Hao LeeYingtao LiuChenkai Dai
Published in: Bioengineering (Basel, Switzerland) (2023)
Cochlear implant (CI) surgery is one of the most utilized treatments for severe hearing loss. However, the effects of a successful scala tympani insertion on the mechanics of hearing are not yet fully understood. This paper presents a finite element (FE) model of the chinchilla inner ear for studying the interrelationship between the mechanical function and the insertion angle of a CI electrode. This FE model includes a three-chambered cochlea and full vestibular system, accomplished using µ-MRI and µ-CT scanning technologies. This model's first application found minimal loss of residual hearing due to insertion angle after CI surgery, and this indicates that it is a reliable and helpful tool for future applications in CI design, surgical planning, and stimuli setup.