Noninvasive Assessment of Mycotic Nail Tissue Using an Ultraviolet Fluorescence Excitation Imaging System.
Vinzent Kevin OrtnerWalfre FrancoMerete HaedersdalPeter Alshede PhilipsenPublished in: Lasers in surgery and medicine (2020)
UV-images clearly depicted all 100 nail samples, with a visibly stronger signal in infected samples. Statistically significant differences (P < 0.05) in signal intensity between mycotic samples and healthy controls were observed for maximum and average fluorescence values. Mean fluorescence values of onychomycotic samples showed 23.9% higher maximum (mycotic: 34.9 AU [standard deviation [SD] 4.7]; healthy: 28.2 AU [SD 1.9]) and 10.2% higher average (mycotic: 27.6 AU [SD 2.0]; healthy: 25.0 AU [SD 0.7]) signal intensity values. Receiver operating characteristic curves demonstrated excellent discriminatory ability (area under the curve > 0.9). Analysis of fluorescence measurements of the reference standard demonstrated very low variation (coefficient of variation = 0.62%) CONCLUSION: Quantitation of u-FEI intensities enables differentiation between healthy and mycotic nail samples, constituting a potential point-of-care tool for cost-effective screening for onychomycosis at a primary care level. Lasers Surg. Med. © 2020 Wiley Periodicals LLC.