Login / Signup

The Anterior Subcutaneous Pelvic Ring Fixator: No Biomechanical Advantages Compared with External Fixation.

Martin Cornelius JordanAnna-Christina BremsTimo HeintelHendrik JansenStefanie DohtRainer H Meffert
Published in: The Journal of bone and joint surgery. American volume (2020)
Although internal fixator devices are placed close to the bone and should therefore maintain greater stiffness, our data did not support the hypothesis of superior stability. Special attention is required when using a curved connecting rod as the rod is easily displaced, placing relevant anatomical structures at risk. These devices require further refinement to avoid potential patient injury.
Keyphrases