Determining spectral response of the National Ignition Facility particle time of flight diagnostic to x rays.

B ReicheltN KabadiJ PearcyM Gatu JohnsonS DannhoffBrandon LahmannJohan FrenjeChikang LiG SutcliffeJ KunimuneRichard PetrassoH SioA S MooreE MariscalE P Hartouni
Published in: The Review of scientific instruments (2023)
The Particle Time of Flight (PTOF) diagnostic is a chemical vapor deposition diamond detector used for measuring multiple nuclear bang times at the National Ignition Facility. Due to the non-trivial, polycrystalline structure of these detectors, individual characterization and measurement are required to interrogate the sensitivity and behavior of charge carriers. In this paper, a process is developed for determining the x-ray sensitivity of PTOF detectors and relating it to the intrinsic properties of the detector. We demonstrate that the diamond sample measured has a significant non-homogeneity in its properties, with the charge collection well described by a linear model ax + b, where a = 0.63 ± 0.16 V -1  mm -1 and b = 0.00 ± 0.04 V -1 . We also use this method to confirm an electron to hole mobility ratio of 1.5 ± 1.0 and an effective bandgap of 1.8 eV rather than the theoretical 5.5 eV, leading to a large sensitivity increase.