IMPACT OF BREATH SAMPLE COLLECTION METHOD AND LENGTH OF STORAGE OF BREATH SAMPLES IN TEDLAR BAGS ON THE LEVEL OF SELECTED VOLATILES ASSESSED USING GAS CHROMATOGRAPHY-ION MOBILITY SPECTROMETRY (GC-IMS).
Barbora CzippelovaSlavomira NovakovaMiroslava SarlinovaEva BaranovicovaAnna UrbanovaZuzana TurianikovaJana Cernanova KrohovaErika HalasovaHenrieta SkovierovaPublished in: Journal of breath research (2024)
The analysis of volatile organic compounds (VOCs) in exhaled air has attracted the interest of the scientific community because it provides the possibility of monitoring physiological and metabolic processes and non-invasive diagnostics of various diseases. However, this method remains underused in clinical practice as well as in research because of the lack of standardized procedures for the collection, storage and transport of breath samples, which would guarantee good reproducibility and comparability of results. The method of sampling, as well as the storage time of the breath samples in the polymer bags used for sample storage and transport, affect the composition and concentration of VOCs present in the breath samples. The aim of our study was to compare breath samples obtained using two methods with fully disposable equipment: Haldane sampling tube intended for direct breath collection and breath samples exhaled into a transparent Tedlar bag. The second task was to monitor the stability of selected compounds of real breath samples stored in a Tedlar bag for 6 hours. Gas chromatography - ion mobility spectrometry (GC-IMS) implemented in the BreathSpec® device was used to analyse exhaled breath. Our results showed a significant difference in the signal intensity of some volatiles when taking a breath sample with a Haldane tube and a Tedlar bag. Due to its endogenous origin, acetone levels were significantly higher when the Haldane tube sampler was used while elevated levels of 2-propanol and unidentified VOC in the Tedlar bag samples likely originated from contamination of the Tedlar bags. The VOC stability study revealed compound-specific signal intensity changes of the selected VOCs with storage time in the Tedlar bags, with some volatiles showing increasing signal intensity during Tedlar bag storage. Our results highlight the importance of careful design and implementation of experiments and clinical protocols to obtain relevant and reliable results.