Characterization of Surgical Tools for Specific Endovascular Navigation.

A BadrouN TardifA EvenP ChaudetN LescanneJ SzewczykA GravouilN HamilaAline Bel-Brunon
Published in: Cardiovascular engineering and technology (2022)
The Young moduli of the shaft and the distal tip of the guidewire were determined. We defined a suitable current intensity to activate the guidewire related to an optimal curvature. Then, the time of activation/deactivation was measured at 1.7 s. On the flip side, parts of the catheters were considered either elastic or viscoelastic. In all cases, the rigidity gradients along the various catheters were highlighted. The characterization of the aforementioned surgical tools provides the opportunity to simulate the endovascular nagivation process.