A proper protocol for routine 18 F-FDG uEXPLORER total-body PET/CT scans.

Huiran HuYanchao HuangHongyan SunKemin ZhouLi JiangJinmei ZhongLi ChenLijuan WangYanjiang HanHubing Wu
Published in: EJNMMI physics (2023)
A 1 min scan with a full dose and a 2 min scan with a half dose are optimal for clinical diagnosis using the HYPER iterative and 2 min and 3 min for OSEM. For quantification of the small lesions, HYPER iterative reconstruction is preferred.
Keyphrases
  • pet ct
  • computed tomography
  • positron emission tomography
  • dual energy
  • image quality
  • clinical practice
  • contrast enhanced