Extreme γ-Ray Radiation Tolerance of Spectrometer-Grade CsPbBr 3 Perovskite Detectors.

Michael C De SienaVladislav V KlepovSergei P StepanoffKhasim Saheb BayikadiLei PanIndra R PandeySujita KarkiDuck Young ChungDouglas E WolfeMercouri G Kanatzidis
Published in: Advanced materials (Deerfield Beach, Fla.) (2023)
The perovskite compound CsPbBr 3 has recently been discovered as a promising room-temperature semiconductor radiation detector, offering an inexpensive and easy-to-manufacture alternative to the current benchmark material Cd 1-x Zn x Te (CZT). We have evaluated the performance of CsPbBr 3 sensors under harsh conditions, such as high radiation doses often found in industrial settings and extreme radiation in space. Results showed minimal degradation in detector performance after exposure to 1 Mrad of Co-60 gamma radiation, with no significant change to energy resolution or hole mobility and lifetime. Additionally, many of the devices were still functional after being exposed to a 10 Mrad dose over three days, and those that did not survive could still be refabricated into working detectors. These results suggest that the failure mode in these devices is likely related to the interface between the electrode and material and their reaction, or the electrode itself and not the material itself. Overall, the study suggests that CsPbBr 3 has high potential as a reliable and efficient radiation detector in various applications, including those involving extreme fluxes and energies of gamma-ray radiation. This article is protected by copyright. All rights reserved.