Positive Reactant Ion Formation in High Kinetic Energy Ion Mobility Spectrometry (HiKE-IMS).
Maria AllersAnsgar T KirkManuel EckermannChristoph SchaeferDuygu ErdogduWalter WissdorfThorsten BenterStefan ZimmermannPublished in: Journal of the American Society for Mass Spectrometry (2020)
In contrast to classical Ion Mobility Spectrometers (IMS) operating at ambient pressure, the High Kinetic Energy Ion Mobility Spectrometer (HiKE-IMS) is operated at reduced pressures of between 10 and 40 mbar and higher reduced electric field strengths of up to 120 Td. Thus, the ion-molecule reactions occurring in the HiKE-IMS can significantly differ from those in classical ambient pressure IMS. In order to predict the ionization pathways of specific analyte molecules, profound knowledge of the reactant ion species generated in HiKE-IMS and their dependence on the ionization conditions is essential. In this work, the formation of positive reactant ions in HiKE-IMS is investigated in detail. On the basis of kinetic and thermodynamic data from the literature, the ion-molecule reactions are kinetically modeled. To verify the model, we present measurements of the reactant ion population and its dependence on the reduced electric field strength, the operating pressure, and the water concentration in the sample gas. All of these parameters significantly affect the reactant ion population formed in HiKE-IMS.