Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study.

Michael NissenSyrine SlimKatharina M JaegerMadeleine FlaucherHanna HuebnerNina DanzbergerPeter Andreas FaschingMatthias W BeckmannStefan GradlBjoern M Eskofier
Published in: JMIR formative research (2022)
Device accuracy varied with activity. Overall, both devices achieved a mean absolute percentage error of just <10%. Thus, they were considered to produce valid results based on the limits established by previous work in the field. Neither device reached sufficient accuracy during seated rest or keyboard typing. Thus, both devices may be eligible for use in respective studies; however, researchers should consider their individual study requirements.