Cochlear implantation programming characteristics and outcomes of cochlear nerve deficiency.
Xingmei WeiSimeng LuBiao ChenJingyuan ChenLifang ZhangYongxin LiYing KongPublished in: European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery (2023)
The CI programming parameters of some CND patients need to be adjusted, and a slower stimulation rate and higher pulse width are required sometimes. CND patients need a higher stimulus level than normal patients but their outcomes are poorer. Non-auditory response should be noticed in CND patients during programming.