Analyzing RPM and RGSC Infrared Camera Sensitivity: A Comparative Study of Breathing Cycle Replication Verifications and Limitations.
Arun BalakrishnanPadmanabhan Ramesh BabuPublished in: Asian Pacific journal of cancer prevention : APJCP (2024)
Data for TrueBeam-STx, Novalis-Tx, and the CT simulator shows discrepancies in replicating the breathing cycle as Seconds Per Breath (SPB) decreases. Effective Infrared (IR) sensitivity is observed until SPB thresholds: 1.8s (TrueBeam-STx), 2.2s (Novalis-Tx), and 2.2s (CT simulator). We should consider values equal to or greater than the mentioned breathing periods. Variations in replicating breathing cycles signal challenges in planning and delivering treatments, especially with lower SPB values. These insights guide clinicians to adapt treatments based on machine-specific capabilities for accurate and reproducible outcomes.