Login / Signup

Organ-Specific Positron Emission Tomography Scanners for Breast Imaging: Comparison between the Performances of Prior and Novel Models.

Yoko SatohKohei HanaokaChihiro IkegawaMasamichi ImaiShota WatanabeDaisuke Morimoto-IshikawaHiroshi OnishiToshikazu ItoYoshifumi KomoikeKazunari Ishii
Published in: Diagnostics (Basel, Switzerland) (2023)
The performances of photomultiplier tube (PMT)-based dedicated breast positron emission tomography (PET) and silicon photomultiplier tube (SiPM)-based time-of-flight (TOF) PET, which is applicable not only to breast imaging but also to head imaging, were compared using a phantom study. A cylindrical phantom containing four spheres (3-10 mm in diameter) filled with 18 F-FDG at two signal-to-background ratios (SBRs), 4:1 and 8:1, was scanned. The phantom images, which were reconstructed using three-dimensional list-mode dynamic row-action maximum likelihood algorithm with various β-values and post-smoothing filters, were visually and quantitatively compared. Visual evaluation showed that the 3 mm sphere was more clearly visualized with higher β and smaller post-filters, while the background was noisier; SiPM-based TOF-PET was superior to PMT-based dbPET in sharpness, smoothness, and detectability, although the background was noisier at the SBR of 8:1. Quantitative evaluation revealed that the detection index (DI) and recovery coefficient (CRC) of SiPM-based TOF-PET images were higher than those of PMT-based PET images, despite a higher background coefficient of variation (CV BG ). The two organ-specific PET systems showed that a 3 mm lesion in the breast could be visualized at the center of the detector, and there was less noise in the SiPM-based TOF-PET image.
Keyphrases