Login / Signup

Tele-robotic distal gastrectomy with lymph node dissection on a cadaver.

Yuma EbiharaSatoshi HiranoYo KurashimaHironobu TakanoKunishige OkamuraSoichi MurakamiToshiaki ShichinoheHajime MorohashiEiji OkiKenichi HakamadaNorihiko IkedaMasaki Mori
Published in: Asian journal of endoscopic surgery (2023)
The purpose of this study is to evaluate the performance of tele-robotic distal gastrectomy (tele-RDG) with lymph node dissection (LND) using a novel Japanese-made surgical robot hinotori™ (Medicaroid, Kobe, Japan) in a cadaver with a presumptive gastric cancer. The Cadaveric Anatomy and Surgical Training Laboratory (CAST-Lab.) at Hokkaido University and Kushiro City General Hospital (KCGH) are connected by a guaranteed type line (1 Gbps), and the distance between the two facilities is 250 km. A patient cart was installed at CAST-Lab, and a surgeon cockpit was installed at KCGH. Tele-RDG with D2 LND was performed on an adult human cadaver. In all surgical processes, the communication environment was stable without image degradation, and the mean round trip time was 40 milliseconds (36.5-55 milliseconds). For tele-RDG with D2 LND, the operation time was 199 minutes without any technical problems. Tele-RDG using hinotori™ was feasible and similar to local robotic RDG.
Keyphrases