Lateral flow immunoassay device to detect staphylococcal enterotoxin B (SEB) in durian candy.

Paquito E ReloxKatherine Ann C IsraelYuki TokuraDaniel Citterio
Published in: Analytical sciences : the international journal of the Japan Society for Analytical Chemistry (2024)
Staphylococcal enterotoxin B (SEB), a potent enterotoxin produced by Staphylococcus aureus, has been implicated in incidences of Staphylococcal food poisoning in the Philippines. The use of lateral flow immunoassay devices to detect this toxin in solid food samples, like durian candy, at the point of sampling is constrained by the requirement for sample purification (e.g. centrifugation). This problem is also true with the other applications of LFIA devices on food samples. To overcome this challenge, a lateral flow immunoassay (LFIA) device capable of detecting SEB in unpurified durian candy sample was developed in this study. A modified LFIA device was assembled with three layers of glass fiber pads functioning as sample pads instead of a conventional cellulose fiber pad. Unlike with the cellulose fiber pad, the glass fiber sample pads acted as filter and allowed the flow of a 1:5 dilution of durian candy. The LFIA device applied to spiked 1:5 diluted durian candy samples achieved a visual limit of detection of 5 ng/mL for SEB, which is twofold lower than reported for previous LFIA devices designed to detect SEB in food samples.