Login / Signup

Transcatheter aortic valve replacement with Evolut R versus Sapien 3 in Japanese patients with a small aortic annulus: The OCEAN-TAVI registry.

Hiromu HaseNobuhiro YoshijimaRyo YanagisawaMakoto TanakaHikaru TsurutaHideyuki ShimizuKeiichi FukudaToru NaganumaKazuki MizutaniMasahiro YamawakiNorio TadaFutoshi YamanakaShinichi ShiraiMinoru TabataHiroshi UenoKensuke TakagiYusuke WatanabeMasanori YamamotoKentaro Hayashidanull null
Published in: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2020)
TAVR for patients with a small annulus using third-generation THVs was associated with high device success. Evolut R seems to be superior to Sapien 3 in hemodynamic performance for patients with a small annulus and body surface area up to 1 year after TAVR. Nevertheless, all-cause mortality at 1 year was similar between both groups.
Keyphrases
  • aortic valve
  • transcatheter aortic valve replacement
  • transcatheter aortic valve implantation
  • aortic stenosis
  • coronary artery disease
  • coronary artery
  • atrial fibrillation
  • pulmonary artery