The OSCAR-MP Consensus Criteria for Quality Assessment of Retinal Optical Coherence Tomography Angiography.
Rebecca WickleinCharmaine YamChristina NollLilian AlyNicolas BanzeEva Feodora RomahnElisabeth WolfBernhard HemmerFrederike Cosima OertelHanna Gwendolyn ZimmermannPhilipp AlbrechtMarius RingelsteinCarmen BaumannNikolaus FeuchtJosef PenkavaJoachim HavlaJonathan A GernertChristian MardinEleni S VasileiouAnneke Van Der WaltOmar Al-LouziSergio CabelloÁngela Vidal-JordanaJulia KrämerHeinz WiendlJana Lizrova PreiningerovaOlga CiccarelliElena Garcia-MartinVeronika KanaPeter A CalabresiFriedemann PaulShiv SaidhaAxel PetzoldAhmed E OthmanBenjamin Kniernull nullPublished in: Neurology(R) neuroimmunology & neuroinflammation (2023)
We propose a 3-step approach for standardized quality control: (1) To define a specific ROI, (2) to assess the occurrence of OCTA artifacts according to the OSCAR-MP criteria, and (3) to evaluate OCTA quality based on the occurrence of different artifacts within the ROI. OSCAR-MP OCTA QC criteria achieved high interrater agreement in an international multicenter study and is a promising QC protocol for application in the context of future clinical trials and studies.