Joint system relaxometry (JSR) and Crámer-Rao lower bound optimization of sequence parameters: A framework for enhanced precision of DESPOT T1 and T2 estimation.
Rui Pedro A G TeixeiraShaihan J MalikJoseph V HajnalPublished in: Magnetic resonance in medicine (2017)
Compared with DESPOT, JSR produces reproducible relaxation maps with improved precision. Further improvement was achieved using CRLB as a protocol design tool. With this combined approach, it is possible to achieve submillimeter maps of ρ,T1,T2, and B0 in an 11-min examination, making the approach appealing for potential clinical use. Magn Reson Med 79:234-245, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.