Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait.
Samuel BladesMatt JensenTrent StellingwerffSandra HundzaMarc D KlimstraPublished in: Sensors (Basel, Switzerland) (2023)
This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, instrumented treadmill (Bertec) during a running protocol. Benchtop testing results showed that both SI and F-Scan had strong positive linearity (Pearson's correlation coefficient, PCC = 0.86-0.97, PCC = 0.87-0.92; RMSE = 15.96 ± 9.49) and mean root mean squared error RMSE (9.17 ± 2.02) compared to the F-Scan on a progressive loading step test. The SI and F-Scan had comparable results for linearity and hysteresis on a sinusoidal loading test (PCC = 0.92-0.99; 5.04 ± 1.41; PCC = 0.94-0.99; 6.15 ± 1.39, respectively). SI had less mean RMSE (6.19 ± 1.38) than the F-Scan (8.66 ±2.31) on the sinusoidal test and less absolute error (4.08 ± 3.26) than the F-Scan (16.38 ± 12.43) on a static test. Both the SI and F-Scan had near-perfect between-day reliability interclass correlation coefficient, ICC = 0.97-1.00) to the F-Scan (ICC = 0.96-1.00). During running, the SI pressure output had a near-perfect linearity and low RMSE compared to the force measurement from the Bertec treadmill. However, the SI pressure output had a mean hysteresis of 7.67% with a 28.47% maximum hysteresis, which may have implications for the accurate quantification of kinetic gait measures during running.