Reading of gafchromic EBT-3 film using an overhead scanner.
H BantanHiroshi YasudaPublished in: Biomedical physics & engineering express (2024)
Gafchromic film, a commercially available radiochromic film, has been developed and widely used as an effective tool for radiation dose verification and quality assurance in radiotherapy. However, the orientation effect in scanning a film remains a concern for practical application in beam profile monitoring. To resolve this issue, the authors introduced a novel method using an overhead scanner (OHS) coupled with a tracing light board instead of a conventional flatbed scanner (FBS) to read Gafchromic EBT3 films. We investigated the orientation effect of the EBT3 film with a regular hexagonal shape after irradiation with 5 Gy x-rays (160 kV, 6.3 mA) and compared the digitized images acquired using a commercially available OHS (CZUR Aura) and a conventional FBS (EPSON GT-X980). As a result, RGB color intensities acquired from the OHS showed significantly lower orientation effect of the color intensities of RGB components than those from FBS. This finding indicates the high potential of the proposed method for achieving more precise two-dimensional dosimetry. Further studies are required to confirm the effectiveness of this method under different irradiation conditions over a wider dose range.
Keyphrases
- room temperature
- reduced graphene oxide
- image quality
- randomized controlled trial
- radiation induced
- early stage
- systematic review
- radiation therapy
- deep learning
- high resolution
- computed tomography
- squamous cell carcinoma
- magnetic resonance
- machine learning
- locally advanced
- convolutional neural network
- rectal cancer
- atomic force microscopy
- dual energy