Preclinical Investigation of Ab Interno Goniotomy Using Three Different Techniques.

David A AmmarEric PorteousMalik Y Kahook
Published in: Clinical ophthalmology (Auckland, N.Z.) (2023)
The various methods evaluated to perform ab interno goniotomy resulted in varying degrees of TM incision or excision. Only the KDB GLIDE device resulted in reliable excision of TM, while the other devices produced incision or minimal excision of tissue with residual leaflets and debris. Use of iAccess resulted in focal disruption of the anterior scleral wall. Because incisional approaches that leave longer residual leaflets may be more prone to fibrosis and closure compared to excisional treatments, clinical correlation will be necessary to better understand the significance of these findings with respect to relative effectiveness of intraocular pressure lowering in eyes with glaucoma.
Keyphrases
  • cataract surgery
  • randomized controlled trial
  • optical coherence tomography
  • laparoscopic surgery
  • stem cells
  • optic nerve
  • surgical site infection