Determination of paraffin oil mist penetration at high flow rates through air-purifying respirators.
Agnieszka BrochockaMałgorzata OkrasaPublished in: International journal of occupational safety and ergonomics : JOSE (2021)
This article presents paraffin oil mist penetration tests of commercially available air-purifying respirators of different construction conducted using the method described by Standard No. ISO 16900-3:2012, which incorporates flow rates (up to 255 l/min) of test aerosol. The testing method reflects differences in work intensity during the use of respirators. Moreover, the experimental stand, designed according to the international specifications, is described. The results show that the higher the paraffin oil mist flow rate, the higher the penetration index, irrespective of the testing method used and the type of respirator investigated. While at high flow rates, filtering half masks of the first protection class (FFP1) met the requirements of their protection class according to European Standard No. EN 149:20001+A1:2009, filtering half masks of the second and the third protection class (FFP2 and FFP3) did not.